Differences in radiation dose for computed tomography of the brain among pediatric patients at the emergency departments: an observational study

Abstract

Background

Computed tomography (CT) is associated with a risk of cancer development. Strategies to reduce radiation doses vary between centers. We compared radiation doses of CT brain studies between pediatric and general emergency departments (EDs), and determine the proportion studies performed within the reference levels recommended by the International Commission on Radiological Protection (ICRP).

Methods

A retrospective review was carried out in a healthcare network consisting of one pediatric ED and three general hospital EDs. Pediatric patients less than 16 years old with CT brain studies performed between 1 January 2015 and 31 December 2018 were included. Information on demographic, diagnosis, volume-averaged computed-tomography dose index and dose length product (DLP) were collected. Effective dose was then calculated from DLP using conversion factors, termed k-coefficients which were derived using a 16 cm head CT dose phantom.

Results

Four hundred and seventy-nine CT brain studies were performed – 379 (79.1%) at the pediatric ED. Seizure (149, 31.1%), head injury (147, 30.7%) and altered mental status (44, 9.2%) were the top three ED diagnoses. The median effective dose estimates were higher in general than pediatric EDs, particularly for those aged > 3 to ≤6 years old [1.57 mSv (IQR 1.42–1.79) versus 1.93 mSv (IQR 1.51–2.28), p = 0.047], > 6 to ≤10 years old [1.43 mSv (IQR 1.27–1.67) versus 1.94 mSv (IQR 1.61–2.59), p = 0.002) and > 10 years old (1.68 mSv (IQR 1.32–1.72) versus 2.03 mSv (IQR 1.58–2.88), p < 0.001). Overall, 233 (48.6%) and 13 (2.7%) studies were within the reference levels recommended by ICRP 60 and 103 respectively.

Conclusions

Radiation doses for CT brain studies were significantly higher at general EDs and less than half of the studies were within the reference levels recommended by ICRP. The development of diagnostic reference levels (DRLs) as a benchmark and clinical justification for performing CT studies can help reduce the radiation risks in the pediatric population.

Peer Review reports

Background

Computed tomography (CT) is a valuable diagnostic tool in the Emergency Department (ED) [1]. However, it involves the use of ionizing radiation at much higher doses as compared to other diagnostic imaging tests such as x-rays [2]. Given the increased utilisation of CT in recent years, concerns have therefore been raised about the potential for developing malignancies resulting from exposure to ionizing radiation during CT studies [1, 3,4,5]. The pediatric population is particularly susceptible due to their radiosensitive organ tissues and long lifespan after radiation exposure, with an increased incidence of radiation-induced cancers such as leukaemia and brain tumours [1, 6,7,8,9,10,11,12,13,14,15].

Previously, radiation doses were similar for pediatric and adult CT techniques – despite the smaller body habitus and increased radiosensitivity of children [16]. Recommendations for reducing ionizing radiation dose for pediatric patients started to emerge in the early 2000s, led by the “As Low As Reasonably Achievable” and “Image Gently” campaigns to raise awareness about the risks associated with CT studies and propose protocols to reduce the radiation dose for children [13, 16,17,18,19]. However, literature on pediatric multi-slice CT studies reported an effective dose of about 4 mSv for CT brain examinations, as opposed to an effective dose of 1 to 2 mSv in adults, suggesting that a gap still exists in clinical practice [20, 21]. In addition, effective doses in pediatric CT head studies can vary across healthcare facilities, particularly between institutions with dedicated pediatric services and those without [22,23,24,25,26,27,28]. Reasons for such disparities include inappropriate clinical justification; patient-related attributes like size, weight, motion, and complexity of medical condition; as well as technical factors like a CT scanner’s acquisition parameters, and the level of training or experience of the CT technologist and radiologist [23, 29]. As such, there is a continual need to push for appropriate justification of pediatric CT studies with optimisation of radiation dose used [30, 31].

To this end, a landmark development was when the American Association of Physicists in Medicine (AAPM) established the size-specific dose estimate (SSDE) which takes into consideration the patient’s size, and allows for optimisation of radiation dose without sacrificing image quality [32,33,34]. Also, the International Commission on Radiological Protection (ICRP) has since updated its age and region-specific effective dose limits from the previous 1990 Publication 60 recommendations to the current 2007 Publication 103, based on the latest available scientific information of the biology and physics of radiation exposure [35, 36]. Finally, the establishment of a CT dose diagnostic reference level (DRL), defined as a radiation dose level to identify situations where the patient dose or administered activity is unusually high, further helps with dose optimisation [37]. In healthcare practices, DRLs help prevent radiation exposures that do not provide additional information for patient clinical management [38]. They also help guide radiology departments review their practices when patient doses significantly deviate from the DRL, and allow comparisons between institutions for quality control purposes. At the time of writing, there is no established national or local DRL guidance for pediatric CT head examinations in Singapore.

Therefore, our primary objective was to compare radiation doses used for pediatric CT brain studies between pediatric and general EDs of a healthcare network in Singapore, and determine the proportion of CT brain studies performed within the reference level recommended by ICRP [35, 36, 39]. (Supplementary Materials). Our secondary objective was to propose a DRL for pediatric CT brain studies performed in the EDs. We hypothesize that for pediatric CT brain studies, the median effective radiation dose administered in general EDs is higher compared to the pediatric ED, and a greater proportion of CT studies is within ICRP 60 and 103 recommendations at pediatric than general EDs. The findings from this work can drive quality improvement initiatives and ultimately, it is about making CT imaging safe for every pediatric patient in the ED.

Methods

Study setting

The public healthcare institutions in Singapore are divided into three healthcare networks, with each serving a specific geographical location. Our healthcare network is the largest and consists of (1) a pediatric tertiary hospital with a pediatric ED and both inpatient and outpatient pediatric services; (2) three adult tertiary hospitals with general EDs and no inpatient or outpatient pediatric service. The pediatric ED is staffed by pediatric emergency physicians while the general EDs are staffed by emergency physicians who have variable experience in pediatric emergency care.

Study design

A retrospective study was carried out involving pediatric patients, defined by age less than 16 years old, who attended the EDs from 1 January 2015 to 31 December 2018 and had a CT brain study performed. Data was collected using a standardized form by assessing the electronic medical records of the ED visit. Information on demographic, diagnosis and radiation dose in terms of volume-averaged computed-tomography dose index (CTDIvol) and dose length product (DLP) were obtained and analysed. All CT brain studies will be analysed, including those repeated in the same ED visit.

This study was approved by Institutional Review Board at SingHealth, with waiver of informed consent.

CT scanner parameters

Across all EDs, CT brain studies were performed using one of the following three CT scanners: Toshiba Aquilion (Toshiba Medical Systems, Tokyo, Japan), Toshiba Aquilion Prime (Toshiba Medical Systems, Tokyo, Japan) or Siemens Somatom Force (Siemens Healthcare, Germany).

Radiation dose parameters

Dose indicators of CTDIvol and DLP were obtained. CTDIvol reflects the mean absorbed radiation within the scan volume based upon standardized CTDI phantoms. For CT brain studies, CTDI quantification is based on a 16 cm diameter Plexiglas phantom [21]. DLP, determined by multiplying the CTDIvol (in mGy) by scan length (in cm), is a measure of a CT scanner’s radiation output/exposure along a patient’s long axis (in mGy-cm) and provides an estimate of the total energy delivered to the CTDI Plexiglas phantom – consequently, the scan length may be estimated by dividing the DLP by the CTDIvol [21].

Effective dose, in mSv, was calculated from DLP by multiplying DLP with age- and region-specific conversion factors, termed k-coefficients (in mSv/(mGy-cm)), recommended by the ICRP 60 and 103 [35, 36, 39]. (Supplementary Materials).

Data analysis

SPSS version 22 (SPSS, Chicago, IL) was used to perform statistical analysis. Frequencies with percentages were used to present categorical data. Mean ± standard deviation (SD) or median (interquartile range, IQR) was used for continuous data depending on normality. Patients were classified into the following five age-group categories – ≤6 months, > 6 months to ≤3 years, > 3 years to ≤6 years, > 6 years to ≤10 years, > 10 years – to compare their effective doses and with the reference levels recommended by ICRP. Chi-square test or Fisher’s exact test was used for association between categorical data. Student’s t-test or Mann-Whitney U test was used for association between continuous data depending on normality. Statistical significance was taken at p less than 0.050.

Results

Demographics

A total of 479 pediatric patients had CT brain studies performed over the study period – 379 (79.1%) in the pediatric ED and 100 (20.9%) in the three general EDs. All of the CT brain studies were non-contrasted. There was no patient with multiple CT brain studies in a single ED visit. The median age of the patients across the four EDs was 7 years (IQR 3 to 12) and there were 290 (60.5%) males. Seizure (149, 31.1%), head injury (147, 30.7%) and altered mental status (44, 9.2%) were the top three ED diagnoses for patients requiring CT brain studies in the EDs (Table 1).

Table 1 Patient Demographics

CTDIvol, DLP and effective dose for CT brain studies

Table 2 shows the parameters of the CT scanners used at the pediatric and general EDs.

Table 2 Parameters of CT Scanners At Pediatric and General EDs

There were significant differences in both median CTDIvol and DLP between the pediatric ED and general EDs in the > 3 to ≤6 years, > 6 to ≤10 years and > 10 years age groups. (Supplementary Materials) The median effective dose for all CT brain studies was 1.68 mSv (IQR 1.43 to 2.10 mSv) which was slightly more than half a year’s worth of background radiation exposure in the United States (3.1 mSv/year) [40]. Across all age groups, there was a trend towards greater median effective doses for CT brain studies performed in the general EDs as compared to the pediatric ED. In particular, significant differences in the CT brain median effective doses were found in the > 3 to ≤6 years (p = 0.047), > 6 to ≤10 years (p = 0.002) and > 10 years (p < 0.001) age groups (Table 3).

Table 3 Comparison of Effective Dose for CT Brain Studies Between Pediatric and General EDs

Proportion of CT brain studies with effective doses within the reference levels recommended by ICRP 60 and 103

Two hundred and thirty-three (48.6%) of all the CT brain studies done across both pediatric and general EDs fell within the reference levels recommended by ICRP 60. The proportion is significantly higher in the pediatric ED where 216 (45.1%) were within ICRP 60 definition, compared to 13 (2.7%) in general EDs (p < 0.001). However, when the newer ICRP 103 definition was used, only 13 (2.7%) of all CT brain imaging studies were within the reference levels, all of which were in the pediatric ED (Table 4).

Table 4 CT Brain Studies with Effective Doses within the Reference Levels Recommended by ICRP 60 and 103

Proposal of a local diagnostic reference level

Based on the recommendations of National Council on Radiation Protection and Measurements (NCRP) Report 172 and ICRP 103, the median CTDIvol and DLP values for each age group may be defined as the local DRL. These proposed DRLs from our study are displayed in Table 5.

Table 5 Proposed Local Diagnostic Reference Level for Pediatric CT Brain Studies

Discussion

CT brain studies remain a sensitive and readily available tool for many diagnostic dilemmas in the ED. CT has the ability to produce images in a quick, non-invasive and reliable manner, making it the best imaging option at the ED over alternative imaging modalities such as paediatric cranial ultrasound or magnetic resonance imaging of the brain. Ultrasound is constrained by the fact that it can only be performed prior to fontanelle fusion in neonates or infants and magnetic resonance imaging requires patients to be absolutely still for a prolonged duration of which the pediatric population may not be able to comply with. CT, however, delivers considerably higher radiation doses than these alternatives. Therefore, minimizing radiation doses during CT brain scans remains an important quality measure in all hospitals.

We found that the effective doses of radiation for CT brain studies were higher in the general EDs than the pediatric ED. By comparing effective doses to the published ICRP 60 and 103 definitions, we found that even in pediatric institutions, less than 50% met the ICRP 60 definitions and less than 5% met the newer ICRP 103 definitions. Therefore, moving forward, we have proposed a local DRL based on the median CTDIvol and DLP values for each age group to improve the safety and quality of CT brain studies performed in pediatric patients at our healthcare network.

The term ‘effective dose’, used by health practitioners worldwide, is regarded as the most appropriate dose descriptor to quantify and communicate the stochastic risks associated with diagnostic procedures involving ionizing radiation [41]. It takes into account the relative sensitivity of a person’s irradiated organs, translating it into a quantifiable estimate of an individual’s biologic detriment (e.g., carcinogenesis) – this is referred to as ‘normalized’ effective dose [35]. In neonates and children, an inherently higher sensitivity to the effects of ionizing radiation can result in doubling of the effective dose delivered to the irradiated anatomic site (e.g., brain) compared to an adult [10, 42].

The age-stratified effective doses of CT brain studies were higher in the younger age groups compared to older age groups; a finding in line with recent publications which supports the inverse relationship between effective dose and age [43, 44]. However, the effective doses for pediatric CT brain studies performed by general EDs were higher, reaching statistical significance in those aged three and above, adding further evidence to the literature that non-dedicated pediatric centers exposed patients to higher radiation doses during CT studies [45,46,47]. In addition, it is concerning that there is a wider variation across effective doses in the general EDs and a higher proportion of studies exceeding the dose recommendations of ICRP 60. As ICRP 103 conversion coefficients are deemed to be more accurate for children of different ages, it is therefore alarming that when benchmarked against the ICRP 103 effective dose recommendations, none of the CT brain studies at the general EDs was within the reference levels [48].

For CT studies, effective dose is directly proportional to the quantity of energy a scanner emits, quantified by CTDIvol and DLP. While DLP values help determine estimations of effective dose using ICRP 103 age-based conversion coefficients, CTDIvol, however, is the CT index that ‘best represents the average dose at a particular point within a scan volume’. [49] The CTDIvol is dependent on several key acquisition parameters such as tube current and scanning rotation time (mAs), tube potential, pitch setting and detector configuration, as well as reconstruction technique and slice thickness; all ultimately contribute to patient’s effective dose [50, 51]. Differences in acquisition parameters of the CT scanners at the EDs may account for the lower doses of radiation at the pediatric ED compared to general EDs. Specifically, general EDs had a higher mAs in all age-groups, pediatric ED had a lower gantry rotation time of 0.28 s, and general EDs had a greater proportion of younger children whose images are acquired at a tube potential of 120 kV voltage setting.

However, acquisition parameters of the CT scanners are not the only reasons for the higher doses used for CT brain studies at the general EDs. While the use of optimal scanning protocols involving pediatric patients may lead to lower radiation doses during CT brain studies, the pediatric protocols are not the default setting at the general EDs, thereby necessitating a switch by the radiographer when the study is being performed [52]. Furthermore, radiologists at the general EDs may be less experienced and familiar with CT studies in pediatric patients, therefore making modifications to protocols in place by increasing doses of radiation to decrease image noise for better diagnostic accuracy [26, 28]. All these deviations can contribute to the higher radiation doses for CT brain studies at the general EDs.

Optimal radiation doses should be established by regulatory authorities as national DRLs using national survey, or healthcare facilities as local DRLs using current practice [39]. However, both national and local DRLs are lacking in Singapore. In our study, we proposed local DRLs based on the median CTDIvol and DLP at pediatric and general EDs. In doing so, we hope to call to attention the need for better regulation of radiation exposure from CT studies in pediatric patients. When compared to European guidelines, the local DRLs were higher, especially those for general EDs. (Supplementary Materials).

This study is an initial but critical step towards understanding where we currently stand in terms of pediatric CT dose levels and how it varies across the EDs in our healthcare network. The processes involved in determining radiation dose estimates, including local DRLs, have provided us with an evaluation framework and tools for optimizing doses for pediatric CT brain studies performed in our EDs. At the same time, they have shown us the potential challenges of establishing consensual DRL for pediatric CT studies. The local DRL values in this review were well below those reported by our American and European counterparts, which is encouraging. However, dose variations found between pediatric and non-pediatric EDs within our healthcare network are troubling. Nonetheless, we hope our findings serve as a stimulus for a concerted nationwide and unified approach to pediatric CT dose optimisation, involving physicians, allied health professionals, and patients.

For institutions, we propose having an active approach to educate and raise awareness among healthcare workers beyond ED and radiology staff as part of our efforts to reduce radiation doses in pediatric patients. Implementation of quality control measures to ensure competence when dealing with pediatric patients and monitoring compliance to departmental pediatric protocols for scanning should be considered, especially in centers without dedicated pediatric services. Finally, it is our responsibility to ensure that pediatric CT studies are clinically indicated, and that their acquisition techniques and protocols are optimized. Frequent reviews of pediatric CT protocols to prevent unnecessary radiation exposure to our younger population from routine day-to-day medical practice are highly encouraged.

Limitations

This study has several limitations. Firstly, this study was conducted in a single healthcare network consisting of a pediatric ED and three general EDs in Singapore. All four centers are tertiary hospitals and academic centers. As EDs in other settings may have different practices, a collaboration involving various institutions from multiple countries would be able to provide a better representation of how radiation doses used for CT studies differed between pediatric and general EDs. Next, we only evaluated non contrasted CT brain studies instead of including all CT studies as the number of CT studies involving contrast and/or other body regions were performed infrequently, leading to inadequate numbers for statistical power and valid comparison of radiation doses used between pediatric and general EDs. We also did not evaluate the indications for CT and the number of CT brain studies with positive or clinically important findings.

In determining the effective dose estimates for pediatric CT brain, we used scanner-derived parameters, as well as age- and tube potential-based conversion coefficients recommended by ICRP 103. CTDIvol provides estimates of radiation output doses based on similar attenuating objects and do not take into account those substantially different in terms of size or shape, particularly in children where significant differences in size is often encountered. Other methods including scanner-specific effective dose estimates and size-specific dose estimates are available. Consequently, patients in this review could have been categorized in other ways such as size (e.g., effective head diameter). Furthermore, we did not evaluate the quality of images as this would have improved our understanding of what constitutes image diagnostic adequacy and acceptability across a spectrum of pediatric radiation dose metrics. Hence, future studies should consider including quantitative and qualitative evaluations of diagnostic image quality for completeness. Lastly, we did not study the long term effects of the higher radiation doses used at the general EDs on the subsequent development of radiation-induced cancers.

Conclusions

CT brain remains an important diagnostic imaging tool that is not easily substituted by alternative imaging modalities. When performed for appropriate indications with optimized technical parameters, the value of the information obtained far exceeds the stochastic risks. However, as general EDs have a tendency to administer higher radiation dose during CT brain study compared to pediatric ED, there must be strategies in place to justify and optimize the use of CT study by adopting imaging protocols with reduced radiation doses based on national or local standards. This will allow for better quality management in pediatric CT imaging, so that radiation exposure can be kept “As Low As Reasonably Achievable” and every pediatric patient can be “Image(d) Gently” in the ED.

Availability of data and materials

The datasets generated and analyzed during the current study are not publicly available as per guidelines by Institutional Review Board at SingHealth, Singapore but may be available from the corresponding author on reasonable request.

Abbreviations

AAPM:

American Association of Physicists in Medicine

CTDIvol :

Volume-averaged computed-tomography dose index

DLP:

Dose-length product

DRL:

Diagnostic reference levels

ED:

Emergency Department

ICRP:

International Commission on Radiological Protection

IQR:

Interquartile range

SD:

Standard deviation

SSDE:

Size-specific dose estimate

References

  1. Brenner DJ, Hall EJ. Computed tomography - an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277–84. https://0-doi-org.brum.beds.ac.uk/10.1056/NEJMra072149.

    Article  CAS  PubMed  Google Scholar 

  2. Rehani MM, Berry M. Radiation doses in computed tomography: the increasing doses of radiation need to be controlled. BMJ. 2000;320(7235):593–4. https://0-doi-org.brum.beds.ac.uk/10.1136/bmj.320.7235.593.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of Cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298(3):317–23. https://0-doi-org.brum.beds.ac.uk/10.1001/jama.298.3.317.

    Article  CAS  PubMed  Google Scholar 

  4. Huda W. Radiation doses and risks in chest computed tomography examinations. Proc Am Thorac Soc. 2007;4(4):316–20. https://0-doi-org.brum.beds.ac.uk/10.1513/pats.200611-172HT.

    Article  PubMed  Google Scholar 

  5. Kharbanda AB, Krause E, Lu Y, Blumberg K. Analysis of radiation dose to pediatric patients during computed tomography examinations. Acad Emerg Med. 2015;22(6):670–5. https://0-doi-org.brum.beds.ac.uk/10.1111/acem.12689.

    Article  PubMed  Google Scholar 

  6. Berrington de Gonzalez A, Salotti JA, McHugh K, Little MP, Harbron RW, Lee C, et al. Relationship between pediatric CT scans and subsequent risk of leukaemia and brain tumours: assessment of the impact of underlying conditions. Br J Cancer. 2016;114(4):388–94. https://0-doi-org.brum.beds.ac.uk/10.1038/bjc.2015.415.

    Article  PubMed  Google Scholar 

  7. Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol. 2001;176(2):289–96. https://0-doi-org.brum.beds.ac.uk/10.2214/ajr.176.2.1760289.

    Article  CAS  PubMed  Google Scholar 

  8. Chida K, Ohno T, Kakizaki S, Takegawa M, Yuuki H, Nakada M, et al. Radiation dose to the pediatric cardiac catheterization and intervention patient. AJR Am J Roentgenol. 2010;195(5):1175–9. https://0-doi-org.brum.beds.ac.uk/10.2214/AJR.10.4466.

    Article  PubMed  Google Scholar 

  9. Frush DP, Donnelly LF, Rosen NS. Computed tomography and radiation risks: what pediatric health care providers should know. Pediatrics. 2003;112(4):951–7. https://0-doi-org.brum.beds.ac.uk/10.1542/peds.112.4.951.

    Article  PubMed  Google Scholar 

  10. Huang WY, Muo CH, Lin CY, Jen YM, Yang MH, Lin JC, et al. Pediatric head CT scan and subsequent risk of malignancy and benign brain tumour: a nation-wide population-based cohort study. Br J Cancer. 2014;110(9):2354–60. https://0-doi-org.brum.beds.ac.uk/10.1038/bjc.2014.103.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Journy N, Rehel JL, Le Ducou Pointe H, Lee C, Brisse H, Chateil JF, et al. Are the studies on cancer risk from CT scans biased by indication? Elements of answer from a large-scale cohort study in France. Br J Cancer. 2015;112(1):185–93. https://0-doi-org.brum.beds.ac.uk/10.1038/bjc.2014.526.

    Article  CAS  PubMed  Google Scholar 

  12. Krille L, Dreger S, Schindel R, Albrecht T, Asmussen M, Barkhausen J, et al. Risk of cancer incidence before the age of 15 years after exposure to ionizing radiation from computed tomography: results from a German cohort study. Radiat Environ Biophys. 2015;54(1):1–12. https://0-doi-org.brum.beds.ac.uk/10.1007/s00411-014-0580-3.

    Article  CAS  PubMed  Google Scholar 

  13. Mathews JD, Forsythe AV, Brady Z, Butler MW, Goergen SK, Byrnes GB, et al. Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ. 2013;346(may21 1):f2360–f.

    Article  Google Scholar 

  14. Pearce MSD, Salotti JAP, Little MPP, McHugh KF, Lee CP, Kim KPP, et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet. 2012;380(9840):499–505. https://0-doi-org.brum.beds.ac.uk/10.1016/S0140-6736(12)60815-0.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Valentin J. International commission on radiological P. the 2007 recommendations of the international commission on radiological protection. Oxford: Published for the International Commission on Radiological Protection by Elsevier; 2007.

    Google Scholar 

  16. Paterson A, Frush DP, Donnelly LF. Helical CT of the body: are settings adjusted for pediatric patients? Am J Roentgenol. 2001;176(2):297–301. https://0-doi-org.brum.beds.ac.uk/10.2214/ajr.176.2.1760297.

    Article  CAS  Google Scholar 

  17. Donnelly LF, Emery KH, Brody AS, Laor T, Gylys-Morin VM, Anton CG, et al. Minimizing radiation dose for pediatric body applications of single-detector helical CT: strategies at a large Children's hospital. AJR Am J Roentgenol. 2001;176(2):303–6. https://0-doi-org.brum.beds.ac.uk/10.2214/ajr.176.2.1760303.

    Article  CAS  PubMed  Google Scholar 

  18. Huda W, Scalzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology. 2000;217(2):430–5. https://0-doi-org.brum.beds.ac.uk/10.1148/radiology.217.2.r00nv35430.

    Article  CAS  PubMed  Google Scholar 

  19. Sodhi KS, Krishna S, Saxena AK, Sinha A, Khandelwal N, Lee EY. Clinical application of ‘justification’ and ‘optimization’ principle of ALARA in pediatric CT imaging: “how many children can be protected from unnecessary radiation?”. Eur J Radiol. 2015;84(9):1752–7. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ejrad.2015.05.030.

    Article  PubMed  Google Scholar 

  20. Brody AS, Frush DP, Huda W, Brent RL. American academy of pediatrics section on R. radiation risk to children from computed tomography. Pediatrics. 2007;120(3):677–82. https://0-doi-org.brum.beds.ac.uk/10.1542/peds.2007-1910.

    Article  PubMed  Google Scholar 

  21. McCollough C, Cody D, Edyvean S, Diagnostic Imaging Council CT Committee. American Association of Physicists in Medicine Report No. 96 - The Measurement, Reporting, and Management of Radiation Dose in CT. 2008.

  22. King MA, Kanal KM, Relyea-Chew A, Bittles M, Vavilala MS, Hollingworth W. Radiation exposure from pediatric head CT: a bi-institutional study. Pediatr Radiol. 2009;39(10):1059–65. https://0-doi-org.brum.beds.ac.uk/10.1007/s00247-009-1327-1.

    Article  PubMed  Google Scholar 

  23. Sadigh G, Kadom N, Karthik P, Sengupta D, Strauss KJ, Frush D, et al. Noncontrast head CT in children: National Variation in radiation dose indices in the United States. AJNR Am J Neuroradiol. 2018;39(8):1400–5. https://0-doi-org.brum.beds.ac.uk/10.3174/ajnr.A5719.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Strauss KJ, Somasundaram E, Sengupta D, Marin JR, Brady SL. Radiation dose for pediatric CT: comparison of pediatric versus adult imaging facilities. Radiology. 2019;291(1):158–67. https://0-doi-org.brum.beds.ac.uk/10.1148/radiol.2019181753.

    Article  PubMed  Google Scholar 

  25. Blumfield E, Zember J, Guelfguat M, Blumfield A, Goldman H. Evaluation of an initiative to reduce radiation exposure from CT to children in a non-pediatric-focused facility. Emerg Radiol. 2015;22(6):631–41. https://0-doi-org.brum.beds.ac.uk/10.1007/s10140-015-1335-4.

    Article  PubMed  Google Scholar 

  26. Goske MJ, Strauss KJ, Coombs LP, Mandel KE, Towbin AJ, Larson DB, et al. Diagnostic reference ranges for pediatric abdominal CT. Radiology. 2013;268(1):208–18. https://0-doi-org.brum.beds.ac.uk/10.1148/radiol.13120730.

    Article  PubMed  Google Scholar 

  27. Nabaweesi R, Ramakrishnaiah RH, Aitken ME, Rettiganti MR, Luo C, Maxson RT, et al. Injured Children Receive Twice the Radiation Dose at Nonpediatric Trauma Centers Compared With Pediatric Trauma Centers. J Am Coll Radiol. 2018;15(1 Pt A):58–64.

    Article  Google Scholar 

  28. Strauss KJ, Goske MJ, Towbin AJ, Sengupta D, Callahan MJ, Darge K, et al. Pediatric chest CT diagnostic reference ranges: development and application. Radiology. 2017;284(1):219–27. https://0-doi-org.brum.beds.ac.uk/10.1148/radiol.2017161530.

    Article  PubMed  Google Scholar 

  29. Boutis K, Thomas KE. Radiation dose awareness and disclosure practice in pediatric emergency medicine: how far have we come? Br J Radiol. 2016;89(1061):20160022. https://0-doi-org.brum.beds.ac.uk/10.1259/bjr.20160022.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Head Injury: Triage, Assessment, Investigation and Early Management of Head Injury in Children, Young People and Adults. National Institute for Health and Clinical Excellence: Guidance. London. 2014.

  31. European Commission. Referral Guidelines for Imaging. Available from http://ec.europa.eu/energy/nuclear/radioprotection/publication/doc/118_en.pdf

  32. Larson DB. Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method. Pediatr Radiol. 2014;44(Suppl 3):501–5. https://0-doi-org.brum.beds.ac.uk/10.1007/s00247-014-3077-y.

    Article  PubMed  Google Scholar 

  33. Seibert JA, Boone JM, Wootton-Gorges SL, Lamba R. Dose is not always what it seems: where very misleading values can result from volume CT dose index and dose length product. J Am Coll Radiol. 2014;11(3):233–7. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jacr.2013.10.010.

    Article  PubMed  Google Scholar 

  34. Yu L, Bruesewitz MR, Thomas KB, Fletcher JG, Kofler JM, McCollough CH. Optimal tube potential for radiation dose reduction in pediatric CT: principles, clinical implementations, and pitfalls. Radiographics. 2011;31(3):835–48. https://0-doi-org.brum.beds.ac.uk/10.1148/rg.313105079.

    Article  PubMed  Google Scholar 

  35. ICRP, 1991. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann ICRP. 1991;21(1–3):4f–201.

  36. ICRP, 2007. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2–4):1–332.

  37. Task Group on Control of Radiation Dose in Computed Tomography. Managing patient dose in computed tomography. A report of the International Commission on Radiological Protection. Ann ICRP. 2000;30(4):7–45.

  38. Clerkin C, Brennan S, Mullaney LM. Establishment of national diagnostic reference levels (DRLs) for radiotherapy localisation computer tomography of the head and neck. Rep Pract Oncol Radiother. 2018;23(5):407–12. https://0-doi-org.brum.beds.ac.uk/10.1016/j.rpor.2018.07.012.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Brady Z, Ramanauskas F, Cain TM, Johnston PN. Assessment of pediatric CT dose indicators for the purpose of optimisation. Br J Radiol. 2012;85(1019):1488–98. https://0-doi-org.brum.beds.ac.uk/10.1259/bjr/28015185.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  40. Metting NF. Ionizing Radiation Dose Ranges (Sievert). 2017. Available at: https://imsva91-ctp.trendmicro.com:443/wis/clicktime/v1/query?url=https%3a%2f%2fwww.pnnl.gov%2fsites%2fdefault%2ffiles%2fmedia%2ffile%2fDose%2520Range%2520Chart%5fSievert%5fJan%25202017%2da.pdf&umid=73FB666D-CA36-D705-A58A-9238160BC2A1&auth=6e3fe59570831a389716849e93b5d483c90c3fe4-b7008eb368e11b2c76590e7e28721ef1bb74df48. Accessed 4 Sept 2021.

  41. McCollough CH, Schueler BA. Calculation of effective dose. Med Phys. 2000;27(5):828–37. https://0-doi-org.brum.beds.ac.uk/10.1118/1.598948.

    Article  CAS  PubMed  Google Scholar 

  42. Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci U S A. 2003;100(24):13761–6. https://0-doi-org.brum.beds.ac.uk/10.1073/pnas.2235592100.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  43. Aw-Zoretic J, Seth D, Katzman G, Sammet S. Estimation of effective dose and lifetime attributable risk from multiple head CT scans in ventriculoperitoneal shunted children. Eur J Radiol. 2014;83(10):1920–4. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ejrad.2014.07.006.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  44. Muhammad NA, Abdul Karim MK, Abu Hassan H, Ahmad Kamarudin M, Ding Wong JH, Ng KH. Diagnostic Reference Level of Radiation Dose and Image Quality among Pediatric CT Examinations in A Tertiary Hospital in Malaysia. Diagnostics (Basel). 2020;10(8):591.

  45. Dougeni E, Faulkner K, Panayiotakis G. A review of patient dose and optimisation methods in adult and pediatric CT scanning. Eur J Radiol. 2012;81(4):e665–83. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ejrad.2011.05.025.

    Article  CAS  PubMed  Google Scholar 

  46. Mohiy HA, Sim J, Seeram E, Annabell N, Geso M, Mandarano G, et al. A dose comparison survey in CT departments of dedicated pediatric hospitals in Australia and Saudi Arabia. World J Radiol. 2012;4(10):431–8. https://0-doi-org.brum.beds.ac.uk/10.4329/wjr.v4.i10.431.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Moss M, McLean D. Pediatric and adult computed tomography practice and patient dose in Australia. Australas Radiol. 2006;50(1):33–40. https://0-doi-org.brum.beds.ac.uk/10.1111/j.1440-1673.2005.01544.x.

    Article  CAS  PubMed  Google Scholar 

  48. ICRP, Khong PL, Ringertz H, Donoghue V, Frush D, Rehani M, et al. ICRP publication 121: radiological protection in pediatric diagnostic and interventional radiology. Ann ICRP. 2013;42(2):1–63. https://0-doi-org.brum.beds.ac.uk/10.1016/j.icrp.2012.10.001.

    Article  Google Scholar 

  49. Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257(1):158–66. https://0-doi-org.brum.beds.ac.uk/10.1148/radiol.10100047.

    Article  PubMed  Google Scholar 

  50. Huda W, Vance A. Patient radiation doses from adult and pediatric CT. AJR Am J Roentgenol. 2007;188(2):540–6. https://0-doi-org.brum.beds.ac.uk/10.2214/AJR.06.0101.

    Article  PubMed  Google Scholar 

  51. Raman SP, Mahesh M, Blasko RV, Fishman EK. CT scan parameters and radiation dose: practical advice for radiologists. J Am Coll Radiol. 2013;10(11):840–6. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jacr.2013.05.032 State of the Art: Iterative CT reconstruction techniques. Radiology. 2015;276(2):339–57, DOI: 10.1148/radiol.2015132766.

    Article  PubMed  Google Scholar 

  52. Bone MC, Schlosser MJ, Perlin JB. Re: injured children receive twice the radiation dose at nonpediatric trauma centers compared with pediatric trauma centers. J Am Coll Radiol. 2018;15(6):825. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jacr.2017.10.036.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was funded by the SingHealth-Duke NUS Emergency Medicine Academic Clinical Program Research Grant (EMRG2018003).

Author information

Authors and Affiliations

Authors

Contributions

TXM performed literature review, collected data, analyzed and interpreted data, as well as drafted and revised the manuscript. MTBMS analyzed and interpreted data, as well as drafted and revised the manuscript. APH and NDBZ collected data and critically reviewed the manuscript. CSL, OYKG, LKP analyzed and interpreted data, as well as critically reviewed the manuscript. PJH conceptualized the study, collected data, analyzed and interpreted data, as well as drafted and revised the manuscript. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Jen Heng Pek.

Ethics declarations

Ethics approval and consent to participate

This study was approved by Institutional Review Board at SingHealth, Singapore (CIRB reference 2019/2360), with waiver of informed consent. All methods were carried out in accordance with relevant guidelines and regulations.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tan, X.M., Shah, M.T.B.M., Chong, SL. et al. Differences in radiation dose for computed tomography of the brain among pediatric patients at the emergency departments: an observational study. BMC Emerg Med 21, 106 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s12873-021-00502-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12873-021-00502-7

Keywords